- Type –?Turbine with light, small head
- Connection –?Using 4-way Unifix quick-connect coupling, rotating 360 degree
- Speed of rotation –?335,000 rpm
- Recommended air pressure –?3.1 bar
- Cooling –?Diffuser, with 3 separate air and water sprays, water antiretraction valve
- Chuck locking mechanism –?Push-button with anti-heating system
- Light transmission –?Multi-strand optical glass conductor
- Lighting –?Halogen
- Cleaning/Sterilisation –?Thermo washer disinfectable / Steam autoclave up to 135 degree C
- Weight –?73 g
- Prestige L ref. –?1600380-001
- Prestige L + 4-way Unifix set ref. –?1700184-001
MAINTENANCE OF TURBINES.
CLEANING
B.REMOVING DIRT/DEPOSITS
- Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
- Dry the nozzles with compressed air then dry the rest of the instrument with a cloth
Do not immerse in an ultrasonic cleaner
C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET
- Shake the?Spraynet?before use
- Use the?Spraynet?attachment?1
(for 2-/3-way and 4-way turbines, see inset below) - Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Spraynet?button down for?1 second
- Clean the exterior of the turbine with a cloth soaked in?Spraynet
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
C’.CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
- Use the?Spraynet?attachment?2
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Spraynet?button down for?1 second
Spray into the opening as shown in the illustrations for ?2-/3-way? and ?4-way?
DISINFECTION
A.DISINFECTION USING SURFACE DISINFECTANT?(follow the manufacturer’s instructions for use)
Do not immerse the instrument in a disinfectant bath
OR
LUBRICATION
OR
B.LUBRICATION WITH LUBRIFLUID
- Shake the?Lubrifluid?before use
- Use the?Lubrifluid?attachment?2,?3?ou?4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Lubrifluid?button down for?1 second
- Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
- Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out
FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
B’.INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY
- Use the?Lubrifluid?attachment?2
- Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Lubrifluid?button down for?1 second
Spray into the opening as shown in the illustrations for ?2-/3-way? and ?4-way?
There are no reviews yet.