The range of Bora turbines combines cutting edge technology with exceptional ergonomics. The advanced performance for which they are renowned has been optimised even further and the purity of their lines make them simply unique. A perfect union of form and function for procedures of unprecedented efficiency and comfort. Bora turbines guarantee impeccable hygiene thanks to the Sealed Head air and water anti-retraction valves and perfect nebulisation provided by the six separate Accu-Spray air and water nozzles. Lastly, they are equipped with a Soft Push button, optimised to provide an easier tool change, and an Accu-Chuck clamp locking mechanism which reduces vibrations and noise. Fitted with an exclusive dual conductor LED system, our turbines provide uniform lighting of the working area, with the intensity of the light almost doubled thanks to brand new technology: hot-pressed multi-strand optical glass.

PREPARATION

Preparation

A.REMOVING THE INSTRUMENT

  • Remove the rotary instrument from the chuck mechanism
Tissue_rubbing

B.PRE-DISINFECTION

  • External disinfection

Do not immerse the instrument in a disinfectant bath.

CLEANING

Cleaning_A

A.CLEAN UNDER RUNNING WATER AT?<38?C

Cleaning_B

B.REMOVING DIRT/DEPOSITS

  • Carefully remove all traces of dirt or deposits from the spray nozzles using the Bien-Air cleaning wire
  • Dry the nozzles with compressed air then dry the rest of the instrument with a cloth

Do not immerse in an ultrasonic cleaner

Cleaning_CSpraynet-Tissue-spray_CTissue_rubbing_C

C.INTERNAL AND EXTERNAL CLEANING WITH SPRAYNET

  • Shake the?Spraynet?before use
  • Use the?Spraynet?attachment?1
    (for 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Spraynet?button down for?1 second
  • Clean the exterior of the turbine with a cloth soaked in?Spraynet

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

Cleaning_C-primeBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

C’.CLEANING THE INSIDE WITH SPRAYNET FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the?Spraynet?attachment?2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Spraynet?button down for?1 second

Spray into the opening as shown in the illustrations for ?2-/3-way? and ?4-way?

DISINFECTION

Disinfectant-Tissue-spray_ATissue_rubbing_A

A.DISINFECTION USING SURFACE DISINFECTANT?(follow the manufacturer’s instructions for use)

Do not immerse the instrument in a disinfectant bath

OR

Thermal-disinfector

B.THERMAL DISINFECTION?(follow the manufacturer’s instructions for use)

LUBRICATION

Lubrification_A1Lubrification_A2

A.LUBRICATION WITH LUBRIMED

OR

Lubrification_B1Support_B

B.LUBRICATION WITH LUBRIFLUID

  • Shake the?Lubrifluid?before use
  • Use the?Lubrifluid?attachment?2,?3?ou?4depending on the type of instrument * (For 2-/3-way and 4-way turbines, see inset below)
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Lubrifluid?button down for?1 second
  • Then, run an operating test for 30 seconds at low speed to remove any surplus lubricant
  • Store immediately: store the instruments with the tip pointing downwards so that any oil residues can flow out

FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

LubrificationBlack-Pearl-Eco_2-3waysBlack-Pearl-Eco_4-ways

B’.INTERNAL LUBRICATION WITH LUBRIFLUID FOR 2-/3-WAY AND 4-WAY TURBINES ONLY

  • Use the?Lubrifluid?attachment?2
  • Keep the tip of the instrument pointing down and protect with an absorbent cloth; press the?Lubrifluid?button down for?1 second

Spray into the opening as shown in the illustrations for ?2-/3-way? and ?4-way?

STERILISATION

Sterilization

A.PLACE THE TURBINE IN A STERILISATION BAG

Sterilizer

B.STERILISE IN ACCORDANCE WITH THE MANUFACTURER’S INSTRUCTIONS

  • Instruments bearing the “135?C” symbol can be sterilised in an autoclave

Only sterilise instruments which have been cleaned and lubricated